Section 8. Operation
Figure 88. Panel-Temperature Gradients (high temperature to low)
Thermocouple Limits of Error
The standard reference that lists thermocouple output voltage as a function of
temperature (reference junction at 0°C) is the NIST (National Institute of
Standards and Technology) Monograph 175 (1993). ANSI (American National
Standards Institute) has established limits of error on thermocouple wire which is
accepted as an industry standard (ANSI MC 96.1, 1975). Table
Limits of Error for
Thermocouple Wire
(p. 331)
gives the ANSI limits of error for standard and special
grade thermocouple wire of the types accommodated by the CR1000.
When both junctions of a thermocouple are at the same temperature, no voltage is
generated, a result of the law of intermediate metals. A consequence of this is that
a thermocouple cannot have an offset error; any deviation from a standard
(assuming the wires are each homogeneous and no secondary junctions exist) is
due to a deviation in slope. In light of this, the fixed temperature-limits of error
(e.g., ±1.0
°
C for type T as opposed to the slope error of 0.75% of the
temperature) in the table above are probably greater than one would experience
when considering temperatures in the environmental range (i.e., the reference
junction, at 0
°
C, is relatively close to the temperature being measured, so the
absolute error — the product of the temperature difference and the slope error —
should be closer to the percentage error than the fixed error). Likewise, because
thermocouple calibration error is a slope error, accuracy can be increased when
the reference junction temperature is close to the measurement temperature. For
the same reason differential temperature measurements, over a small temperature
gradient, can be extremely accurate.
To quantitatively evaluate thermocouple error when the reference junction is not
fixed at 0°C limits of error for the Seebeck coefficient (slope of thermocouple
voltage vs. temperature curve) are needed for the various thermocouples. Lacking
this information, a reasonable approach is to apply the percentage errors, with
perhaps 0.25% added on, to the difference in temperature being measured by the
330
Содержание CR1000
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 32: ......
Страница 36: ......
Страница 38: ......
Страница 40: ......
Страница 60: ...Section 4 System Quickstart Figure 16 PC200W View Line Graph 60 ...
Страница 96: ......
Страница 98: ...98 ...
Страница 302: ......
Страница 350: ...Section 8 Operation Figure 91 Pulse Sensor Output Signal Types Figure 92 Switch Closure Pulse Sensor 350 ...
Страница 453: ...Section 8 Operation Figure 115 Using the Keyboard Display 453 ...
Страница 454: ...Section 8 Operation 8 8 1 Data Display Figure 116 Displaying Data with the Keyboard Display 454 ...
Страница 456: ...Section 8 Operation Figure 118 Real Time Custom 456 ...
Страница 457: ...Section 8 Operation 8 8 1 3 Final Memory Tables Figure 119 Final Memory Tables 457 ...
Страница 458: ...Section 8 Operation 8 8 2 Run Stop Program Figure 120 Run Stop Program 458 ...
Страница 460: ...Section 8 Operation Figure 122 File Edit 460 ...
Страница 461: ...Section 8 Operation 8 8 4 PCCard Memory Card Display Figure 123 PCCard CF Card Display 461 ...
Страница 478: ......
Страница 506: ......
Страница 536: ......
Страница 636: ......
Страница 642: ......
Страница 644: ......
Страница 676: ......
Страница 677: ......