Section 7. Installation
used to access them.
•
No single interface accesses all Status/Settings/DTI. Interfaces used for
access include the following:
Table 6.
Status/Setting/DTI: Access Points
Access Point
Locate in...
Settings Editor
Device Configuration Utility
,
LoggerNet Connect screen
,
PakBus Graph.
See
Datalogger Support Software — Details
(p.
Status
View as a data table in a
numeric monitor
(p. 521).
DataTableInfo
View as a data table in a
numeric monitor
(p. 521).
Station Status
Menu item in
datalogger support software
(p. 654).
Edit Settings
Menu item in
PakBusGraph
software.
Settings
Menu item in CR1000KD Keyboard Display
Configure,
Settings
status.keyword/settings.keyword
Syntax in CRBasic program
1
Information presented in
Station Status
is not updated automatically. Click the
Refresh
button to update.
Note
Communication and processor bandwidth are consumed when generating
the
Status
and
DataTableInfo
tables. If the CR1000 is very tight on processing
time, as may occur in very long or complex operations, retrieving the
Status
table
repeatedly may cause
skipped scans
(p. 487).
603/Settings/DTI
(p. 603)c
an be set or accessed using CRBasic instructions
SetStatus()
or
SetSetting()
.
For example, to set the setting
StationName
to
BlackIceCouloir
, the following
syntax is used:
SetSetting("StationName","BlackIceCouloir")
where
StationName
is the keyword for the setting, and
BlackIceCouloir
is the set
value.
Settings can be requested by the CRBasic program using the following syntax:
x = Status.[setting]
where
Setting
is the keyword for a setting.
For example, to acquire the value set in setting
StationName
, use the following
statement:
x = Status.StationName
7.7.1.4 Configuration with Executable CPU: Files
Many CR1000 settings can be changed remotely over a telecommunication link
either directly, or as discussed in section
Configuration with CRBasic Program
(p.
as part of the CRBasic program. These conveniences come with the risk of
inadvertently changing settings and disabling communications. Such an
occurence will likely require an on-site visit to correct the problem if at least one
of the provisions discussed in this section is not put in place. For example,
115
Содержание CR1000
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 32: ......
Страница 36: ......
Страница 38: ......
Страница 40: ......
Страница 60: ...Section 4 System Quickstart Figure 16 PC200W View Line Graph 60 ...
Страница 96: ......
Страница 98: ...98 ...
Страница 302: ......
Страница 350: ...Section 8 Operation Figure 91 Pulse Sensor Output Signal Types Figure 92 Switch Closure Pulse Sensor 350 ...
Страница 453: ...Section 8 Operation Figure 115 Using the Keyboard Display 453 ...
Страница 454: ...Section 8 Operation 8 8 1 Data Display Figure 116 Displaying Data with the Keyboard Display 454 ...
Страница 456: ...Section 8 Operation Figure 118 Real Time Custom 456 ...
Страница 457: ...Section 8 Operation 8 8 1 3 Final Memory Tables Figure 119 Final Memory Tables 457 ...
Страница 458: ...Section 8 Operation 8 8 2 Run Stop Program Figure 120 Run Stop Program 458 ...
Страница 460: ...Section 8 Operation Figure 122 File Edit 460 ...
Страница 461: ...Section 8 Operation 8 8 4 PCCard Memory Card Display Figure 123 PCCard CF Card Display 461 ...
Страница 478: ......
Страница 506: ......
Страница 536: ......
Страница 636: ......
Страница 642: ......
Страница 644: ......
Страница 676: ......
Страница 677: ......