INSTRUCTIONAL MANUAL
2750L-P/4500L-P/6000L-P
August, 2001
Rev. K
-76-
Table 3-17
NON-CNF STATUS MESSAGES
CONDITION DEVICE
RESPONSE
Loaded output
Fault
F07ACS01 <SP> (DEV): <SP>OUTPUT
CH01/02/03 VOLT FAULT <CR><LF>
Source Temperature Fault
F07ACS01 <SP> (DEV): <SP> SOURCE
CH01/02/03 TEMPERATURE <CR><LF>
Program out of range for VOLT,
FREQ, PANG or CURL
F07ACS01 <SP> (MOD): <SP> VOLT/FREQ/PANG/CURL
RANGE ERROR <CR> <LF>
Unrecognizable OP-code or
missing OP-code
F07ACS01 <SP> (MOD): <SP>SYNTAX ERROR <CR><LF>
Good Status
<SP><CR><LF>
3.8.4 NOUN
MODIFIERS
Noun modifiers are a CIIL mnemonic indicating the characteristic of the signal to be described.
3.8.4.1 VOLT
The VOLT (voltage) noun modifier may be contained in a string as an instruction to the power source with SET, SRX or
SRN. Depending on the OP-code, an output voltage (or voltages) may be programmed or set as an upper or lower limit
in 0.1-volt increments up to the range limit. The channel indicator which precedes the modifier determines which phase
(or phases) will be so set. The number which follows the modifier in the string determines the amplitude to be
programmed. For example, to set the output voltage of all three phases to 110 volts, the following string is transmitted:
FNC ACS :CH00 SET VOLT 110 <CR> <LF>
The noun modifier may also be used in a string that uses the sensor (measurement) capability of the power source.
An example of requesting the measurement of power source phase A output voltage follows. Note that the noun
modifier precedes the channel indicator.
FNC ACS VOLT :CH01 <CR> <LF>
FTH VOLT <CR> <LF>
When the output voltage of phase A is 110.0 the power source responds with:
<SP> 110.0 <CR> <LF>
3.8.4.2 FREQ
The FREQ (frequency) noun modifier may be contained in a string as an instruction to the power source with SET, SRX
or SRN. The channel indicator which precedes the modifier determines which phase (or phases) will be so set. Any one
of the four channel indicators may be used in the string. The number which follows the modifier in the string determines
the frequency to be programmed. For example, to program all phases to 440 Hertz, the following string is transmitted:
FNC ACS :CH00 SET FREQ 440 <CR><LF>
Содержание 12000L-1P
Страница 27: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 23 Figure 2 3 FUNCTION TEST SETUP...
Страница 98: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 94 Figure 4 3 FLOW DIAGRAM A REMOTE CALIBRATION...
Страница 99: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 95 Figure 4 3 FLOW CHART continued...
Страница 108: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 104 This page left intentionally blank...
Страница 110: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 106 Figure 5 1 AC POWER SYSTEM BLOCK DIAGRAM...
Страница 114: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 110 This page left intentionally blank...
Страница 119: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 115 Figure 6 1 MODULE LOCATION...
Страница 120: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 116 This page left intentionally blank...
Страница 133: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 129 Figure 1 STEADY STATE VOLTAGE...
Страница 135: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 131 Figure 8 9 FREQUENCY TRANSIENT...
Страница 136: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 132 Figure 10 11 ABNORMAL OPERATION...
Страница 137: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 133 Figure 12 13 ABNORMAL FREQUENCY...
Страница 138: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 134 Figure 14 EMERGENCY VOLTAGE...
Страница 140: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 136 This page left intentionally blank...
Страница 148: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 144 FIGURE 2...