INSTRUCTIONAL MANUAL
2750L-P/4500L-P/6000L-P
August, 2001
Rev. K
-4-
Table 1-2
SPECIFICATIONS 4500L-P / 2750L-P / 6000L-P
(All specifications apply using external sense, 23 ±5
°
C, constant line and load conditions unless specified otherwise,
after 30 minute warm-up.)
ELECTRICAL
OUTPUT SPECIFICATIONS
VA
CURRENT PER PHASE
(At full output voltage)
VOLTAGE RANGE
MODEL TOTAL PER
PHASE 135
270
50
°
C 35
°
C 50
°
C 35
°
C 50
°
C 35
°
C 50
°
C 35
°
C
2750L-3
2700 3000 900 1000 6.66 7.40 3.30 3.70
2750L-1 2750
3000
2700
3000
20.0
22.22
10.00
11.11
4500L-3 4500
5000
(1)
1500
1667
(1)
11.12 12.34
(1)
5.56 6.17
(1)
4500L-1 4500
5000
(1)
4500 5000
(1)
33.34 37.04
(1)
16.67 18.52
(1)
6000L-3 5330
6000
1770
2000
13.12
14.82
6.56
7.41
6000L-1
5330 6000 5330 6000 39.35 44.44 19.68 22.22
9000L/2-3 9000
10000
(1)
3000 3333
(1)
22.24 24.68
(1)
11.12 12.34
(1)
9000L/2-1 9000
10000
(1)
9000 10000
(1)
66.66 74.08
(1)
33.33 37.04
(1)
12000L/2-3
10600 12000 3540 4000 26.24 29.62 13.12 14.81
12000L/2-1 10600
12000
10600
12000
78.7
88.88
39.35
44.44
13500L/3-3 13500
15000
(1)
4500 5000
(1)
33.34 37.04
(1)
16.67 18.52
(1)
13500L/3-1 13500
15000
(1)
13500 15000
(1)
100.0 111.2
(1)
50.0 55.6
(1)
18000L/3-3
15900 18000 5300 6000 39.35 44.44 19.68 22.22
18000L/3-1 15900
18000
15900
18000
118.0
133.4
59.0
66.7
27000L/6-3 27000
30000
(1)
9000 10000
(1)
66.66 74.08
(1)
33.33 37.04
(1)
27000L/6-1 27000
30000
(1)
27000 30000
(1)
200.0 222.2
(1)
100.0 111.1
(1)
NOTE: Increased current at 35
°
C is available up to 2000 Hz or 440Hz for 6000L.
(1)
For UP option output VA and Current is limited to 50°C values.
For other Voltage Ranges:
Current per phase = VA (per phase)/Voltage Range
Power factor:
0 to unity at full rated output VA
Peak Repetitive Current:
2.5 times the maximum TRMS current at 50
°
C
2750L
&
6000L:4
times
TRMS
current
at
50
°
C
Peak Non-repetitive Current:
(10ms, 20ms for 6000L) 2.7 times the maximum TRMS current at 50
°
C
2750L & 6000L: 4 times TRMS current at 50
°
C
Содержание 12000L-1P
Страница 27: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 23 Figure 2 3 FUNCTION TEST SETUP...
Страница 98: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 94 Figure 4 3 FLOW DIAGRAM A REMOTE CALIBRATION...
Страница 99: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 95 Figure 4 3 FLOW CHART continued...
Страница 108: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 104 This page left intentionally blank...
Страница 110: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 106 Figure 5 1 AC POWER SYSTEM BLOCK DIAGRAM...
Страница 114: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 110 This page left intentionally blank...
Страница 119: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 115 Figure 6 1 MODULE LOCATION...
Страница 120: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 116 This page left intentionally blank...
Страница 133: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 129 Figure 1 STEADY STATE VOLTAGE...
Страница 135: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 131 Figure 8 9 FREQUENCY TRANSIENT...
Страница 136: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 132 Figure 10 11 ABNORMAL OPERATION...
Страница 137: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 133 Figure 12 13 ABNORMAL FREQUENCY...
Страница 138: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 134 Figure 14 EMERGENCY VOLTAGE...
Страница 140: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 136 This page left intentionally blank...
Страница 148: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 144 FIGURE 2...