INSTRUCTIONAL MANUAL
2750L-P/4500L-P/6000L-P
August, 2001
Rev. K
-120-
8.3.4 EMERGENCY
OPERATION
(Refer to MIL-704D Doc. 5.2.5)
1. Voltage per Figure 14, page 156
2. Frequency per Figure 15, page 156
8.4
KEYPAD ENTRY (Refer to page 31 for Keyboard flow chart)
To perform a test from the key board, the following key sequence is required:
704
ENT
The following screen will appear:
MIL704D:SelA,B,C
ENT=all
CLR=EXIT
Pressing the A, B, C or any combination selects the phase in test. Press ENT without the phase select for
simultaneous three-phase test.
The following screen appears for a short time.
TEST
A,B,C
CLR
to
Reselect
The next screen is:
Apply
Nom
Output
Press
ENT
When ENT is selected the following screen appears:
Press MNU to
Select
Test
The MNU screen has two lines of selection shown at a time.
There are 3 different types of operations that can be selected from a MENU acreen. If the word MENU appears for the
item selected, another MENU screen will be displayed. If the word TEST appears fir the item selected, the test will
start. The display will return to the previous screen if the word RETURN appears for the item selected.
The Main Menu will appear as follows:
1=Steady
St
Menu
2=Transient
Menu
3=Abnormal
Menu
4=Emergency
Menu
5=MIL704D
Test
6=Return
If key 1 is selected "Steady State: from the Main Menu, the following Menu will appear:
1=Voltage
Test
2=Unbalance
Test
3=Phase
dif
Test
4=Wave
dist
Test
Содержание 12000L-1P
Страница 27: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 23 Figure 2 3 FUNCTION TEST SETUP...
Страница 98: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 94 Figure 4 3 FLOW DIAGRAM A REMOTE CALIBRATION...
Страница 99: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 95 Figure 4 3 FLOW CHART continued...
Страница 108: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 104 This page left intentionally blank...
Страница 110: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 106 Figure 5 1 AC POWER SYSTEM BLOCK DIAGRAM...
Страница 114: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 110 This page left intentionally blank...
Страница 119: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 115 Figure 6 1 MODULE LOCATION...
Страница 120: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 116 This page left intentionally blank...
Страница 133: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 129 Figure 1 STEADY STATE VOLTAGE...
Страница 135: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 131 Figure 8 9 FREQUENCY TRANSIENT...
Страница 136: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 132 Figure 10 11 ABNORMAL OPERATION...
Страница 137: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 133 Figure 12 13 ABNORMAL FREQUENCY...
Страница 138: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 134 Figure 14 EMERGENCY VOLTAGE...
Страница 140: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 136 This page left intentionally blank...
Страница 148: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 144 FIGURE 2...