Troubleshooting
49
3.
Press
Apply
to confirm selection and restart the Control Unit (see Figure 42).
Note:
Pressing
Cancel
returns to the previous menu and displays the list of
available languages.
Figure 42. Language Confirmation Screen
Troubleshooting
The Control Unit provides system errors, channel errors and channel messages to
communicate status information. Descriptions of these errors/messages and
troubleshooting steps are provided in this section.
System Errors
A system error affects both channels and stops ultrasound therapy on both channels.
System errors are related to system failures, control unit temperature conditions and
device cross connection errors. When a system error occurs, an error message with an
alert icon displays and an audible tone sounds. Read the message and follow the
instructions to attempt to correct the error. Press the
Audio Pause
button on the
Control Unit to silence the alert tone for 5 minutes. When the tone is silenced the alert
icon on the display is crossed out. If the issue cannot be resolved by following the
onscreen instructions, note the error code in parentheses at the end of the message
and contact your local EKOS representative or call +1
.
425
.
415.3100.
Note
: In the event that the Control Unit is not operational and ultrasound therapy has
stopped, infusion of the solutions will not be interrupted.
See Figure 43 and Figure 44 for examples of system error messages: