Ultrasound Operation
32
7.
Press the
Start
button on the Control Unit for the appropriate channel,
when prompted (see Figure 22).
WARNING
!
Never transmit output power (ultrasound therapy ON) from
the Control Unit to the EkoSonic™ Device unless it is placed
within the patient anatomy, solution is running through the
drug lumen, and coolant is flowing through the coolant
lumen.
Always turn OFF the ultrasound before removing
the Ultrasonic Core from the Infusion Catheter
,
otherwise, overheating of the device may occur, potentially
causing damage to the device, burns and/or interruption of
therapy. If the device is damaged in this manner and then
used to continue therapy, vascular injury or tissue burns
could occur.
Note:
If the
Start
button is not pressed within 5 minutes of successfully
connecting the device, the reminder alert tone activates, the alert icon
appears on the display, and the “Ready” title bar flashes alternating between
white and yellow.
Pressing the
Audio Pause
button
on the Control Unit silences the
reminder for 5 minutes.
Figure 22. Reminder to Start Treatment Screen
Monitoring Ultrasound
1.
Confirm ultrasound therapy is running or ON.