Safety Information
14
infusion stand adhere to following guidelines (see Attaching the Control
Unit to an Infusion Stand Section):
•
Always use an infusion stand that meets the specifications in
these Instructions for Use.
•
Follow the mounting instructions described in these Instructions
for Use to ensure the Control Unit is securely attached to the
infusion stand.
•
Attach the Control Unit to a position on the infusion stand such
that the combined weight of the infusion pumps and Control Unit
is as low as practical.
•
When transporting the infusion stand or CU 4.0 Cart, hold the
pole in the middle or use the handle if available to ensure the
best control.
Failure to follow these guidelines could result in the infusion stand
tipping or equipment falling, resulting in bodily injury.
WARNING
!
Always use the handle when carrying the Control Unit. Failure to do so
could cause the Control Unit to be dropped resulting in bodily injury.
WARNING
!
Before transporting the patient or moving the Control Unit, unplug the
power cord from the wall outlet and secure the Connector Interface
Cable near the patient. Pulling of the EkoSonic™ Device may cause
the device to dislodge, loss of access and patient injury.
WARNING
!
The Control Unit user interface presents a historical graph of the
ultrasound power delivered. The power data should not be used for
diagnostic purposes. Diagnosis of drug/ultrasound therapy should
always be made using other clinical diagnostic tools (e.g., fluoroscopy).
An incorrect diagnosis could result in prolonged therapy or inadequate
therapy.
WARNING
!
Thoroughly clean the Control Unit and Connector Interface Cable after
each use. When cleaning, adhere to the following guidelines (See
Cleaning the Control Unit Section):
•
Follow all cleaning instructions provided in these Instructions for
Use.
•
Turn off the Control Unit and unplug it from the power source
before cleaning.
•
Only use the recommended cleaning agents.
•
Do not spray cleaners directly onto the Control Unit or Connector
Interface Cable.
•
Do not expose the Control Unit or the Connector Interface Cable
to autoclaving or other sterilizers.
Failure to follow these cleaning guidelines could result in infection to the
patient or equipment damage.
Содержание EKOS
Страница 65: ...Symbols and Indicators 61 ...