Troubleshooting
51
Message
Description
Error
Code
Possible Cause
Troubleshooting Step
3. When the message disappears,
resume ultrasound.
Control Unit is too
cold.
E005
Control Unit internal
temperature has
fallen below the
operating
temperature.
1. Place the Control Unit in a warmer
environment.
2. Wait for the Control Unit to warm to
room temperature.
3. When the message disappears,
resume ultrasound.
Device cables are
crossed.
E006
Device cables are
cross
‐
connected
between two
different Connector
Interface Cables.
1. Disconnect all device connectors
and reconnect, ensuring that the
Infusion Catheter and the
Ultrasonic Core from a single
EkoSonic™ Device are connected
to one Connector Interface Cable.
2. Press
Start
to begin ultrasound.
Channel Errors
A channel error affects one specific channel and stops ultrasound therapy to the device
associated with that channel. The channel that does not have a channel error will not
be interrupted or affected by the error. Channel errors are related to channel errors,
device temperature errors, cable errors and connection errors. Channel errors can
occur on both channels at the same time. When a channel error occurs, read the
message and follow the instructions to attempt to correct the error. If the issue cannot
be resolved by following the onscreen instructions, note the error code in parentheses
at the end of the message and contact your local EKOS representative or call
+1
.
425
.
415.3100.
Note
: In the event that a specific channel is not operating and ultrasound therapy
cannot be delivered, infusion of the solutions will not be interrupted.
See Figure 45 and Figure 46 below for examples of channel error messages: