Safety Information
13
is damaged in this manner and then used to continue therapy, vascular
injury or tissue burns could occur.
WARNING
!
Avoid spilling fluids on the cables and Control Unit. Do not immerse any
portion of the cables or Control Unit in water or other fluids. Fluids may
cause the equipment to perform inaccurately or fail, which could result in
vascular injury. Additionally, fluids could result in fire or electric shock.
WARNING
!
The Control Unit is intended for use with the EkoSonic™ Device,
approved cables and electronic devices. Do not connect non-EKOS
devices, cables (e.g., power cord) or electronic equipment to the Control
Unit. Use of non-approved devices, cables and electronic equipment
could result in electric shock or burn, increased electromagnetic
interference (EMI) emissions, decreased immunity to EMI from other
sources, or delay of therapy.
WARNING
!
The reusable Connector Interface Cable and Control Unit are NOT
STERILE. Use caution if maintaining a sterile operating field after
connecting to the EkoSonic™ Device. Failure to do so could result in
patient infection.
WARNING
!
Inspect all connectors, cables and the Control Unit for damage prior to
use. If damaged (e.g., exposed wires), do not use and replace
immediately.
WARNING
!
Ensure all connectors are dry before connecting the Connector Interface
Cable to the Control Unit or the Connector Interface Cable to the
EkoSonic™ Device. If connectors do become wet, thoroughly dry prior
to use. Fluid in the connectors may affect sensor accuracy and result in
vascular injury.
WARNING
!
Connect system cables only to their proper connectors as marked and
ensure connections are secure. Failure to do so could result in electric
shock or burn.
WARNING
!
Secure device cables and the Connector Interface Cable using standard
practices to preserve the integrity of the device, minimize device
movement at the insertion site, and prevent device dislodgement. If the
Connector Interface Cable is dropped, it has sufficient weight to pull on
the device and possibly dislodge it from where it has been inserted into
the patient.
WARNING
!
The Control Unit should be placed on a flat table or attached to the
EKOS™ CU 4.0 Cart or an
infusion stand during use. If attached to an