Power
22
Note:
In the event the time elapses and the Control Unit automatically shuts down, the
ultrasound runtime will be saved. Infusion of the solutions will not be interrupted since
the solutions are delivered by a separate infusion pump. Once the Control Unit is
plugged into AC power, turned back ON and ultrasound is started, the runtime will
resume for each channel where it left off.
Battery Error
During operation on AC power, if the internal battery cannot provide power to the
Control Unit, a battery error message with an alert icon displays. The red battery gauge
with an X indicates the battery is defective. Even with a battery error, the Control Unit
can continue to operate as long as it is connected to AC power. Do not attempt to open
the Control Unit and change the battery. Contact your EKOS representative for
information on replacing the battery.
The battery error notification is dismissed by pressing the
Dismiss
button on the touch
screen. The battery error notification appears whenever the Control Unit is connected
to AC power, is turned ON and a battery error is detected (see Figure 10).
If a battery error occurs when the Control Unit is not plugged into AC power, the Control
Unit may not power up when the power button is pressed. If the Control Unit does
power up, a system error message will display indicating the Control Unit will shut down
in three minutes.
If the Control Unit is turned ON when the battery error occurs, the Control Unit will
immediately turn off. To bypass these battery errors, plug the Control Unit into AC
power, power ON and operate normally.
Note
: In the event that the battery is fully depleted or a battery error occurs and the
Control Unit automatically turns OFF, infusion of the solutions will not be interrupted.