Safety Information
16
Caution
When using two EkoSonic™ Devices (one in channel A and one in
channel B), care should be taken to ensure that desired actions (e.g.,
start ultrasound, stop ultrasound, connect cables, disconnect cables,
respond to an alert message) are performed on the correct channel(s) or
device(s).
Caution
When using two EkoSonic™ Devices (one in channel A and one in
channel B), care should be taken to not cross the device cables. Ensure
each pair of device cables is connected to a single Connector Interface
Cable. If the device cables are crossed, ultrasound is not operational
and could result in delay of ultrasound therapy.
Caution
Store the Control Unit in an area of good ventilation and under specified
conditions. (See Specifications Section). Failure to store the Control
Unit as specified could result in equipment failure leading to delay of
ultrasound therapy.
Caution
This equipment has been tested and found to comply with the limits for
medical devices in BS EN 60601
‐
1
‐
2. These limits are designed to
provide reasonable protection against harmful electromagnetic
interference in a typical medical installation. However, portable and
mobile RF communications equipment can affect medical electrical
equipment. Use of the Control Unit with a power cord or accessory other
than as specified in this Instructions for Use may result in increased
electromagnetic interference (EMI) emissions or decreased immunity to
EMI from other sources. (See Specifications Section)
This equipment generates, uses, and can radiate energy and, if not
installed and used in accordance with the instructions, may cause
harmful interference to other devices in the vicinity or be affected by
interference from other devices. There is no guarantee that interference
will not occur in a particular installation. If this equipment does cause
harmful interference to other devices, which can be determined by
turning this equipment off and on, the user is encouraged to try to correct
the interference by one or more of the following measures:
•
Reorient or relocate the receiving device.
•
Increase the separation between the equipment.
•
Connect this equipment into an outlet on a circuit different from
that to which the other device(s) are connected.
•
Consult the manufacturer for help.
If the Control Unit must be used directly adjacent to other equipment,
observe the function of the Control Unit to verify normal operation in that
configuration. If this equipment is adversely affected by interference