Pull the knob to the open position in order to
open the dewar evacuation valve and evacu-
ate the dewar.
10
• If the desired vacuum has been achieved,
gently push the knob to the closed position
in order to close the dewar evacuation valve.
• Purge the section between vacuum pump
and vacuum adapter.
11
Rotate the knob several turns counterclock-
wise in order to screw the union nut off the
dewar evacuation valve. Be careful not to
purge the dewar.
12
Pull the knob to the open position.
13
• Loosen the coupling nut and remove the
vacuum adapter from the connecting piece
of the dewar.
• Reinstall the MCT detector into the spec-
trometer (chapter 5.10)
14
Check whether a signal is detected and the optics works properly. More
details are described in chapter 5.6.
Table 6.16: Evacuating detector
Содержание IFS 125M
Страница 1: ...IFS 125M User Manual 1015996 ...
Страница 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Страница 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Страница 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Страница 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Страница 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Страница 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Страница 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Страница 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Страница 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Страница 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Страница 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...
Страница 227: ...227 Bruker Optik GmbH IFS 125M User Manual ...
Страница 228: ...228 IFS 125M User Manual Bruker Optik GmbH ...