48
IFS 125M User Manual
Bruker Optik GmbH
Operation 5
5.6
Checking the signal
Especially after you have replaced a spectrometer component (source, laser, beamsplit-
ter, detector), it is advisable to check whether a signal is detected, and to check the sig-
nal intensity (signal amplitude).
If a liquid nitrogen-cooled detector is used, the detector must be cooled (chapter 5.9)
before a signal can be detected.
5.6.1
Procedure
1.
Make sure that the spectrometer sample compartment is empty, or contains a gas
cell the windows of which are transparent in the spectral range desired.
2.
On the OPUS
Measure
menu, select the
Advance Measurement
command.
3.
Select appropriate measurement parameters.
4.
Click the
Check Signal
tab.
5.
Activate the
Interferogram
option button.
Figure 5.1:
OPUS - Check Signal tab
B
A
Содержание IFS 125M
Страница 1: ...IFS 125M User Manual 1015996 ...
Страница 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Страница 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Страница 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Страница 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Страница 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Страница 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Страница 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Страница 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Страница 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Страница 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Страница 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...
Страница 227: ...227 Bruker Optik GmbH IFS 125M User Manual ...
Страница 228: ...228 IFS 125M User Manual Bruker Optik GmbH ...