6.7.3
Removing beamsplitter
Remove the beamsplitter to prevent it from breaking while demounting the scanning
arm, or during transport.
The handling instructions are described in chapter 5.8.1.
6
To fix the small retaining clip of the clamp
make sure that the clip is exactly positioned
below the boreholes (see arrow), located on
clamp.
➣
The top of the scanner and the clamp
must be in touch, or must have a very
small gap in between.
Fasten the clip screws.
Table 6.21: Transporting spectrometer - Moving and fixing scanner
1
Turn the main power switch in the OFF posi-
tion to switch off the spectrometer.
2
• Unlock the release lever, move it further up.
• Hold the beamsplitter - with two hands - on
the knobs.
• Carefully remove the beamsplitter straight
upwards without catching an edge.
Table 6.22: Transporting spectrometer - Removing beamsplitter
Содержание IFS 125M
Страница 1: ...IFS 125M User Manual 1015996 ...
Страница 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Страница 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Страница 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Страница 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Страница 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Страница 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Страница 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Страница 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Страница 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Страница 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Страница 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...
Страница 227: ...227 Bruker Optik GmbH IFS 125M User Manual ...
Страница 228: ...228 IFS 125M User Manual Bruker Optik GmbH ...