146
IFS 125M User Manual
Bruker Optik GmbH
Web Interface 8
If you have already sent a command to the optics, one or two additional buttons includ-
ing the corresponding drop-down lists (or entry line) are displayed. See the following fig-
ure.
A) This drop-down list below the first entry line includes the command(s) previously
sent to the spectrometer. To repeat a previous command select it from the drop-
down list and click the
Previous commands
button.
B) The
Values for LSR
button is only displayed if the previous command contains the
Syntax Type
drop-down list. The
Values for...
drop-down list includes all possible
values for a certain command. To set a value select the value from the drop-down
list and click once the
Values for...
button.
C) After the command has been sent to the spectrometer, all occurring warnings and
errors are displayed in the
Messages
group field. In our example, the
LSR=0
com-
mand has been sent. This command causes a HeNe laser to switch off which in turn
deactivates the scanner control. This state of the optics results in the
HeNe-Laser
is
off
or
No laser signal
error message.
8.3.1.1
Error messages
An error message contains the following information:
• Creation time of the message
• Unique error message code (see also chapter 8.3.1.2)
• Importance level of an error/warning message indicated by a letter, e.g. Warning,
Answer, Fatal, Informational, Static (see also chapter 8.3.1.2)
• Origin of the message (device or routine name)
• The command that caused the error message. If it is a general message, this cell
is blank.
• Error message text
i
If you do not know which values are accepted by a certain command (e.g.
APT
), send
the command without any value as a dummy command to the spectrometer. As a result,
the
Values for...
drop-down list including all possible values is displayed. Moreover, the
error message
Value not in device table
will be displayed, which you can ignore.
Figure 8.5:
Command list
A
B
C
Содержание IFS 125M
Страница 1: ...IFS 125M User Manual 1015996 ...
Страница 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Страница 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Страница 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Страница 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Страница 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Страница 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Страница 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Страница 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Страница 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Страница 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Страница 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...
Страница 227: ...227 Bruker Optik GmbH IFS 125M User Manual ...
Страница 228: ...228 IFS 125M User Manual Bruker Optik GmbH ...