Bruker Optik GmbH
IFS 125M User Manual
Maintenance 6
6.3
Replacing source
6.3.1
Source types
The spectrometer can be equipped with the following source types:
Standard source type:
MIR source
Where is the source
located?
On the rear spectrometer side.
When has the source to be
replaced?
Only if it is defective.
How is it indicated that the
source is defective?
The status light in the lower right end of
the OPUS interface becomes red. The
error message
Source is broken or not
connected
is displayed.
Tools required:
None
Handling instructions:
The source becomes hot during spectro-
meter operation. Avoid any skin contact.
Has the source to be
aligned?
All sources used are pre-aligned, and do
not have to be aligned after maintenance.
Table 6.6: Source features
Source type
Estimated lifetime
(hours)
Part number
MIR source, globar, 12V
9000
Q328/7
NIR/VIS source, tungsten, 12V
9000
Q428/7
Blackbody source, emitting area:
100mm x 100mm
S310
Table 6.7: Source types
Содержание IFS 125M
Страница 1: ...IFS 125M User Manual 1015996 ...
Страница 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Страница 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Страница 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Страница 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Страница 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Страница 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Страница 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Страница 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Страница 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Страница 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Страница 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...
Страница 227: ...227 Bruker Optik GmbH IFS 125M User Manual ...
Страница 228: ...228 IFS 125M User Manual Bruker Optik GmbH ...