Maintenance 6
6.2.3
Signal decrease over optical path
The decrease of the peak-to-peak signal amplitude over the full optical path should be
less than 35%.
1
Connect a dual channel oscillograph
a
to the
TEST adapter located on the mobile electron-
ics unit.
Pin assignment of TEST adapter:
1 LASA (Laser A)
2 AGND (Ground for 1 + 3)
3 LASB (Laser B)
4 FWDLSW
5 DGND
6 BWDLSW
7 LASXAF
8 DGND
9 SCM+ (Scanner Voice Coil)
10 SCM- (Scanner Voice Coil)
11 DCANA
12 - 16 Spare
➣
Do NOT short-circuit any pins!
2
Connect one end of the laser signal test cable
to the
Test
connector located on the electron-
ics unit.
3
Connect the other end of the laser signal test cable to the oscillograph.
4
• On the
OPUS
Measure menu, select the
Optic Setup and Service
com-
mand.
• Click the
Control Panel
tab
b
.
Table 6.2: Signal decrease over optical path
Содержание IFS 125M
Страница 1: ...IFS 125M User Manual 1015996 ...
Страница 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Страница 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Страница 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Страница 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Страница 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Страница 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Страница 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Страница 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Страница 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Страница 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Страница 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...
Страница 227: ...227 Bruker Optik GmbH IFS 125M User Manual ...
Страница 228: ...228 IFS 125M User Manual Bruker Optik GmbH ...