161
Bruker Optik GmbH
IFS 125M User Manual
Web Interface 8
ABP Absolute Peak
Position:
Enter the desired absolute peak position manually in the
corresponding entry field. This position determines the
default scanning position of the interferometer.
For double-sided data acquisition, the scanner moves
around this position. This command is used to set the peak
position which is displayed on the
Check Signal
tab of the
Measurement
dialog.
• Sending a positive value saves the absolute peak position
permanently.
• Sending a negative value sets the absolute peak position
only temporarily, i.e. the peak position is not saved. After
restarting the spectrometer, the permanently saved value
is used.
• Sending the
ABP=0
command uses the peak position
value of the last measurement. This value is saved per-
manently.
• Sending the
ABP=-1
command restores the saved peak
position. This command cancels the temporarily peak
position that has been set by sending a negative value.
For additional information see
.
Command
Definition
Table 8.13: Direct commands
Содержание IFS 125M
Страница 1: ...IFS 125M User Manual 1015996 ...
Страница 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Страница 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Страница 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Страница 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Страница 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Страница 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Страница 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Страница 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Страница 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Страница 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Страница 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...
Страница 227: ...227 Bruker Optik GmbH IFS 125M User Manual ...
Страница 228: ...228 IFS 125M User Manual Bruker Optik GmbH ...