54
IFS 125M User Manual
Bruker Optik GmbH
Operation 5
i
If the beamsplitter has been changed, the peak position changes and must therefore be
saved. Select the
Advanced Measurement
command of OPUS and click the
Check Sig-
nal
tab. Then, click the
Save Peak Position
button.
If you use VIS and UV beamsplitters, the interferometer has to be realigned by the
OPUS
Auto Align
function as follows:
1.
Check signal intensity (chapter 5.6).
2.
Once an interferogram is displayed, click the
Auto Align
button.
➣
If you check
Store
mode and activated the
Spectrum
option button, you can
observe the progress visually.
➣
The alignment is most sensitive at high wavenumbers and large apertures. For
example, to align the VIS beamsplitter use a blue or green optical filter and an
aperture of 1.85 mm. Make sure that the detector is not overloaded.
➣
Never use the
Auto Align
function when no (or only a very small) interferogram is
displayed or when working at low wavenumbers. This may lead to severe mis-
alignment.
3
• Unlock the release lever, move it further up.
• Hold the beamsplitter - with two hands - on
the knobs.
• Carefully remove the beamsplitter straight
upwards without catching an edge.
4
• Hold the beamsplitter - with two hands - on
the knobs and insert the new beamsplitter,
with the electrical contacts facing to the
scanner side. The two guiding pins have to
engage into the slots located on the block.
• Push down the beamsplitter completely, until
you feel resistance.
• Lock the release lever, move it further down.
5
The spectrometer beeps when it detects the new beamsplitter and automati-
cally starts scanning.
• Put on the lid again.
• Use the cross-head screwdriver. Push the screws downwards and rotate
them clockwise.
• Check whether a signal is detected, and the optics works properly. More
details are described in chapter 5.6.
Table 5.5: Changing beamsplitter
Содержание IFS 125M
Страница 1: ...IFS 125M User Manual 1015996 ...
Страница 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Страница 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Страница 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Страница 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Страница 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Страница 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Страница 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Страница 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Страница 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Страница 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Страница 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...
Страница 227: ...227 Bruker Optik GmbH IFS 125M User Manual ...
Страница 228: ...228 IFS 125M User Manual Bruker Optik GmbH ...