Maintenance 6
2
Loosen the fixing screw counterclockwise to
remove it from its stand-by position.
3
Rotate the fixing screw clockwise into the fix-
ture on the scanner.
4
When the screw is fixed, the outer scanner
cannot be moved any more.
5
To fix the inner scanner use the supplied
clamp.
➣
To position the clamp make sure that the
clamp screws fit into the boreholes (see
arrow), located on the inner scanner
housing.
➣
Fasten the clamp screws (see arrows).
Pay attention, as the clamp touches the
top side of the movable structure, but
does not press it down! This avoids defor-
mation of the spring leaves.
Table 6.21: Transporting spectrometer - Moving and fixing scanner
Содержание IFS 125M
Страница 1: ...IFS 125M User Manual 1015996 ...
Страница 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Страница 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Страница 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Страница 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Страница 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Страница 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Страница 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Страница 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Страница 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Страница 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Страница 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...
Страница 227: ...227 Bruker Optik GmbH IFS 125M User Manual ...
Страница 228: ...228 IFS 125M User Manual Bruker Optik GmbH ...