55
Bruker Optik GmbH
IFS 125M User Manual
Operation 5
5.9
Cooling detector
As the operating temperature of the MCT detector used in the spectrometer is signifi-
cantly below room temperature, the detector is filled with liquid nitrogen to be able to
achieve the required operating temperature.
The MCT detector used inside the spectrometer has to be cooled in regular time inter-
vals.
Hold time of MCT detector:
Typically 16 hours
Delivery content of detector:
• Funnel
• Protective dewar plug
Cooling medium required:
• Liquid nitrogen
• About 1l/filling (16 hours)
Indications of a weakening cooling
effect:
• Decreased signal intensity
• No signal at all
• OPUS status light becomes red
• Instrument status message in OPUS
reads:
Detector not ready
.
Table 5.6: Detector features
CAUTION
Improper use of liquid nitrogen:
Can cause frostbites, damage to eye tissue, and asphyxiation.
➣
Avoid any skin contact.
➣
Always wear face shields or safety goggles.
➣
Use liquid nitrogen only in well-ventilated rooms.
➣
Observe the safety instructions described in chapter 5.9.1.
Содержание IFS 125M
Страница 1: ...IFS 125M User Manual 1015996 ...
Страница 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Страница 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Страница 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Страница 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Страница 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Страница 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Страница 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Страница 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Страница 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Страница 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Страница 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...
Страница 227: ...227 Bruker Optik GmbH IFS 125M User Manual ...
Страница 228: ...228 IFS 125M User Manual Bruker Optik GmbH ...