58
IFS 125M User Manual
Bruker Optik GmbH
Operation 5
5.10
Changing detector
The basic spectrometer configuration is equipped with a liquid-nitrogen-cooled MCT
detector, mounted on a dovetail guide.
5.10.1
Changing procedure
When changing the detector you have to mind the order described in the following.
Removing covers
1
Removing scanning arm cover:
• Open the clips on each side.
• Put your hand below the clip and push the
clip upwards.
• Pull the clip towards you until you feel resis-
tance.
i
All the clips have to be in their outermost
position, otherwise you get problems when
lifting off the cover.
2
• Two persons are required to lift off the cover
of the long scanning arm, one at the front
and one at the back end.
• Put your hands below the hangers.
• Carefully lift off the cover from the scanning
arm.
Cover not lift off properly.
Risk of scanner damage.
➣
Lift off the cover straight and far
upwards.
Table 5.9: Changing detector - Removing covers
NOTE
Содержание IFS 125M
Страница 1: ...IFS 125M User Manual 1015996 ...
Страница 8: ...8 IFS 125M User Manual Bruker Optik GmbH Table of Contents ...
Страница 14: ...14 IFS 125M User Manual Bruker Optik GmbH Introduction 1 ...
Страница 30: ...30 IFS 125M User Manual Bruker Optik GmbH Installation 3 ...
Страница 66: ...66 IFS 125M User Manual Bruker Optik GmbH Operation 5 ...
Страница 170: ...170 IFS 125M User Manual Bruker Optik GmbH Web Interface 8 ...
Страница 196: ...196 IFS 125M User Manual Bruker Optik GmbH Connection Ports 10 ...
Страница 202: ...202 IFS 125M User Manual Bruker Optik GmbH Spare Parts B ...
Страница 208: ...208 IFS 125M User Manual Bruker Optik GmbH Dimensional Drawings D ...
Страница 214: ...214 IFS 125M User Manual Bruker Optik GmbH Firmware Update E ...
Страница 222: ...222 IFS 125M User Manual Bruker Optik GmbH Addresses G ...
Страница 226: ...226 IFS 125M User Manual Bruker Optik GmbH Index ...
Страница 227: ...227 Bruker Optik GmbH IFS 125M User Manual ...
Страница 228: ...228 IFS 125M User Manual Bruker Optik GmbH ...