Symptom
Check
Actions
There are inlet pressure surges. Is the heater input incorrect ?
Check and rectify if necessary.
Is the fluid outgassing ?
Inspect the condition of the pump
fluid. Refill with new fluid if necessary.
Is there a leak in the vacuum system
ahead of the EO inlet ?
Check and seal any leak found.
Is there a trapped volume in the
vacuum system ?
Check and rectify, if necessary.
Have ‘O’ ring seals been excessively
lubricated ?
Check and rectify, if necessary.
Is there an excessive fluid charge in
the EO pump ?
Check the level of fluid in the pump.
Drain the fluid if necessary.
Table 15 – Fault finding (Sheet 2 of 2)
Содержание EO2K
Страница 1: ...B304 09 880 Issue G Original Instruction Manual EO2K EO4K and EO6K Diffusion Pumps ...
Страница 2: ...This page intentionally blank ...
Страница 6: ...EO2K EO4K and EO6K Diffusion Pumps PAGE iv Issue G iv Aug 06 This page intentionally blank ...
Страница 14: ...EO2K EO4K and EO6K Diffusion Pumps PAGE 8 Issue G 8 Aug 06 This page intentionally blank ...
Страница 20: ...EO2K EO4K and EO6K Diffusion Pumps TECHNICAL DATA PAGE 14 Issue G 14 Aug 06 Figure 10 EO6K dimensions mm ...
Страница 22: ...EO2K EO4K and EO6K Diffusion Pumps PAGE 16 Issue G 16 Aug 06 This page intentionally blank ...
Страница 36: ...EO2K EO4K and EO6K Diffusion Pumps PAGE 30 Issue G 30 Aug 06 This page intentionally blank ...
Страница 52: ...EO2K EO4K and EO6K Diffusion Pumps PAGE 46 Issue G 46 Aug 06 This page intentionally blank ...
Страница 54: ...EO2K EO4K and EO6K Diffusion Pumps PAGE 48 Issue G 48 Aug 06 This page intentionally blank ...
Страница 58: ...EO2K EO4K and EO6K Diffusion Pumps PAGE 52 Issue G 52 Aug 06 This page intentionally blank ...
Страница 61: ...This page intentionally blank ...