Guidance and manufacturer’s declaration – electromagnetic immunity
The SpiroCare is intended for use in the electromagnetic environment specified below.
The customer or the user of the SpiroCare should assure that it is used in such an environment.
IMMUNITY
test
IEC 60601
test level
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-
2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-
4
± 2 kV for power
supply lines
±
1 kV for
input/output lines
± 2 kV for power supply
lines
±
1 kV for input/output
lines
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-
5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should
be that of a typical
commercial or hospital
environment.
Voltage dips,
short
interruptions
and
voltage
variations
on power
supply
input lines
IEC 61000-4-
11
<5%UT
(>95%dip in UT)
for 0,5 cycle
40 %UT
(60 %dip in UT)
for 5, 6 cycles
70 %UT
(30 %dip in UT)
for 25,30 cycles
<5%UT
(>95%dip in UT)
for 5 s
<5%UT
(>95%dip in UT)
for 0,5 cycle
40 %UT
(60 %dip in UT)
for 5, 6 cycles
70 %UT
(30 %dip in UT)
for 25,30 cycles
<5%UT
(>95%dip in UT)
for 5 s
Mains power quality should
be that of a typical
commercial or hospital
environment.
If the user of the SpiroCare
requires continued
operation during power
mains interruptions,
it is recommended that the
SpiroCare be powered
from an uninterruptible
power supply or a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-
8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.
Содержание SpiroCare
Страница 1: ...SpiroCare Operation Manual Ver 1 1 2017 04 14 www ebionet com...
Страница 54: ...54 135 SCOMBA 1 1 SpiroCare Operation Manual FVC Test Pint Form...
Страница 57: ...57 135 SCOMBA 1 1 SpiroCare Operation Manual FVC Test Print Form...
Страница 62: ...62 135 SCOMBA 1 1 SpiroCare Operation Manual SVC Test Print Form...
Страница 67: ...67 135 SCOMBA 1 1 SpiroCare Operation Manual MVV Test Print Form...
Страница 72: ...72 135 SCOMBA 1 1 SpiroCare Operation Manual Calibration Print Form...
Страница 100: ...100 135 SCOMBA 1 1 SpiroCare Operation Manual Data Erase All data in Files and Worklist will be deleted...