69/135
SCOMBA-1.1
SpiroCare Operation Manual
Calibration should be conducted in the following order: Violation of order might cause
error in measurement.
1) Connect conductor of measurement part of spirometer to the main body.
2) Connect the mouthpiece installed in the spirometer handle to mouth of syringe.
Insert it tightly to prevent any leakages.
3) Press start button. If the machine is ready for the calibration, it beeps.
4) Move syringe handle back and forth ten times each.
Make sure to move the handle after you hear the machine beep.
5) Decide whether to apply the calibrated setting to the device.