A graph as the visual representation of flow and volume will be drawn along the
movement of the syringe. The results of previous volume measurement and error (%) will
be displayed on the screen as shown below if the syringe makes the round trip 5 times.
Press the “Exit” button and return to the initial calibration menu to start again if no beep
sound is heard each time.
Select ‘Print’ from the menu to output Calibration result.
Select ‘Accept' from the menu to apply Calibration result to the device.
Select 'Exit' from the menu to end Calibration and exit to the top menu.
Note
-
It is recommended that the revision of the lung capacity measuring instrument
should be practiced each quarter and if the temperature and atmosphere
greatly changes the test should be essentially practiced.
-
Re-practice the revision if your revision result does not belong to +/- 3% and
redo the test until your revision result belongs to +/-3%.
-
Calibrate the spirometer by changing the speed of the flow of the calibration
syringe (0.5~12 L/Sec). Repeat the calibration until the calibrated value reaches
the tolerance range (2.91~3.09L) when 3L calibration syringe is used.
Содержание SpiroCare
Страница 1: ...SpiroCare Operation Manual Ver 1 1 2017 04 14 www ebionet com...
Страница 54: ...54 135 SCOMBA 1 1 SpiroCare Operation Manual FVC Test Pint Form...
Страница 57: ...57 135 SCOMBA 1 1 SpiroCare Operation Manual FVC Test Print Form...
Страница 62: ...62 135 SCOMBA 1 1 SpiroCare Operation Manual SVC Test Print Form...
Страница 67: ...67 135 SCOMBA 1 1 SpiroCare Operation Manual MVV Test Print Form...
Страница 72: ...72 135 SCOMBA 1 1 SpiroCare Operation Manual Calibration Print Form...
Страница 100: ...100 135 SCOMBA 1 1 SpiroCare Operation Manual Data Erase All data in Files and Worklist will be deleted...