Power
▣
AC Power
When AC Power is connected to the device, power LED light will be green; if the battery
is installed, charging will commence.
<AC Power> <Battery Power>
▣
Battery Power
The device will get its power from the battery and the unit will display the battery power
icon (shown above), when the system is powered on, the battery is connected to the
device and the AC power is disconnected. When the battery power is low, an alarm
sound will ring from the device speaker and LCD display will show a “Battery Low”
message. Connect the AC power immediately or the device will automatically shut
down in a 1minutes.
•
Time to full recharge from full discharge: Max. 3 hours
•
Duration of continuous usage after full recharge: if running ECG record in 12
channel format by choosing to set successive function for about 360 min. at
25mm/s and 10mm/mV or in the absence of ECG record, it is possible to record a
max. of 200 ECGs.
▣
Display Battery Power Status
: Battery Fully Charged
: Battery Charge Half-Full
: Battery Charge Low
: Battery Almost Fully Drained
▣
Replacing Battery
When replacing battery for this device, the same type of battery should be used.
•
Type: Lithium-ion battery 3ICR19/65 ( 11.1V-2600mAh )
•
When to replace: Battery will automatically be charged when the device is
connected to AC power, and cannot be charged when separated from the
device. Battery is designed to have a charging cycle of 300 times or more. If the
device only lasts 20 minutes or less on battery power, the battery needs to be
replaced. Additionally, when a battery pack is damaged or leaking chemicals,
replace it immediately. Do not use damaged battery packs with the device.
WARNING
EXPLOSION HAZARD —
DO NOT incinerate the battery or store at high temperatures. Serious injury or death
could result.
Содержание SpiroCare
Страница 1: ...SpiroCare Operation Manual Ver 1 1 2017 04 14 www ebionet com...
Страница 54: ...54 135 SCOMBA 1 1 SpiroCare Operation Manual FVC Test Pint Form...
Страница 57: ...57 135 SCOMBA 1 1 SpiroCare Operation Manual FVC Test Print Form...
Страница 62: ...62 135 SCOMBA 1 1 SpiroCare Operation Manual SVC Test Print Form...
Страница 67: ...67 135 SCOMBA 1 1 SpiroCare Operation Manual MVV Test Print Form...
Страница 72: ...72 135 SCOMBA 1 1 SpiroCare Operation Manual Calibration Print Form...
Страница 100: ...100 135 SCOMBA 1 1 SpiroCare Operation Manual Data Erase All data in Files and Worklist will be deleted...