125/135
SCOMBA-1.1
SpiroCare Operation Manual
System messages
The following system messages may occur while you are operating this system.
You may be required to perform some action. If you perform the recommended
actions and the condition still remains, contact authorized service personnel.
No. System message
Cause
Solution
1
Please check
date and time
Internal coin battery is
discharged.
Please contact CS (Customer
Support) team of Bionet to
replace old coin type battery with
new one
2
Address is not
valid
When wrong network
address(IP, SM, GW) is
entered
Check the right network address
before input
3
Invalid value.
[Use: 0~9, A~F]
When wrong character is
entered at time of
inputting Mac address
Please input valid character such
as 0~9, A to F in number and
alphabet respectively
4
There is no patient
ID. Please enter
patient ID.
When you implement
AUTO or NET button
without patient ID
Please enter patient ID in patient
information window to implement
5
Wrong password.
Please re-enter
your password.
When wrong password is
entered
Check and input right password
6
Enter password
length error.
Please re-enter
your password
When wrong number of
characters is entered as
password
Please check and input valid and
correct password
7
The internal
memory is full.
Please erase
other files or
transfer to
external storage.
When over 200 files are
stored at internal
memory
Erase files in File menu
8
Not enough free
disk space
USB memory device has
insufficient space to
receive files.
Remove unnecessary files from
USB memory (Minimum 500KB
space is needed to export 1 file)
9
Please check the
USB memory.
When USB memory stick is
not inserted, or is out of
order
Please check and insert USB
memory stick in right position.
Note that USB memory file format
supports FAT, FAT32 only (this can
be checked on PC)
Содержание SpiroCare
Страница 1: ...SpiroCare Operation Manual Ver 1 1 2017 04 14 www ebionet com...
Страница 54: ...54 135 SCOMBA 1 1 SpiroCare Operation Manual FVC Test Pint Form...
Страница 57: ...57 135 SCOMBA 1 1 SpiroCare Operation Manual FVC Test Print Form...
Страница 62: ...62 135 SCOMBA 1 1 SpiroCare Operation Manual SVC Test Print Form...
Страница 67: ...67 135 SCOMBA 1 1 SpiroCare Operation Manual MVV Test Print Form...
Страница 72: ...72 135 SCOMBA 1 1 SpiroCare Operation Manual Calibration Print Form...
Страница 100: ...100 135 SCOMBA 1 1 SpiroCare Operation Manual Data Erase All data in Files and Worklist will be deleted...