127/135
SCOMBA-1.1
SpiroCare Operation Manual
22
Fail to send Image
When file transmission to
PACS store server has
failed
Retry after checking network
conditions
23
Fail to connect
Store Server
When there is
Connection failure
between PACS
store server and device
Check network condition with
PACS store server
24
Transmission failed
file is existing.
Are you sure to
delete?
When the transmission
failed file to PACS store
server needs to be
deleted
Delete the failed file after
transmitting the file to PACS store
server
25 Recharge
Battery!
When the battery level is
Low
Please connect power cable or
charge battery
26 Printer:
Time
out.
Printer stops working due
to technical problem of
device
React the printer
27
Printer: Busy
continue
When the wrong printer
data were input
React the printer
28
Printer:
Temperature high
When the Printer gets
heated due to long and
continuous use of printer
Please stop the use of printer for
at least one minute and try again
29
Printer: Paper
empty
When the printing paper
is completely out
Please put new roll of printing
paper for use
30
Printer: Printer not
ready
When there is no ‘ready’
answer from printer
module
React the printer
31
Printer: Unknown
data
When the wrong
protocol was put into
the printer
React the printer
32 Abnormal
End!
When the printer was
abnormally forced to
stop during printing
Re-start printing
33 Please check
handle!
- When the Spirometer
handle is not connected
- When the spirometer
handle is disconnected
during test
Check the Spirometer handle is
connected
34
Correction
factor(s) outside
of acceptable
range of 0.8~1.2
When the factor value is
out of tolerance range
during the calibration
Repeat the calibration until the
calibrated values are within the
tolerance range
35
Please repeat the
calibration until
the calibrated
value reaches the
tolerance range
3% (2.91~3.09L)!
When the difference
values are out of
tolerance range during
the calibration
Repeat the calibration until the
calibrated values are within the
tolerance range
Содержание SpiroCare
Страница 1: ...SpiroCare Operation Manual Ver 1 1 2017 04 14 www ebionet com...
Страница 54: ...54 135 SCOMBA 1 1 SpiroCare Operation Manual FVC Test Pint Form...
Страница 57: ...57 135 SCOMBA 1 1 SpiroCare Operation Manual FVC Test Print Form...
Страница 62: ...62 135 SCOMBA 1 1 SpiroCare Operation Manual SVC Test Print Form...
Страница 67: ...67 135 SCOMBA 1 1 SpiroCare Operation Manual MVV Test Print Form...
Страница 72: ...72 135 SCOMBA 1 1 SpiroCare Operation Manual Calibration Print Form...
Страница 100: ...100 135 SCOMBA 1 1 SpiroCare Operation Manual Data Erase All data in Files and Worklist will be deleted...