SpiroCare Operation Manual
5) System Installation
Precautions For Installation
While installing SpiroCare, please pay attention to following items:
- Use the equipment between the ambient temperature 5 to 40 (
℃
25 to 104 ) and
℉
humidity 10 -90
%
.
- Check the power cord is properly connected, and the probe carefully handled.
- Do not plug multiple cords in a power outlet.
- Install and operate SpiroCare unit on a flat surface.
- If you experience noise, ground the device.
- Do not use a power cord that may make a connection noise.
- Device settings will be recorded in the internal memory even when it is off.
- Prevent any shock or excessive force that may cause damage to the device.
- Place the device away from any dust or flammable materials.
Power Connection
The equipment needs electrical power to operate. Plug in one end of the power cable
to wall socket and the other to SpiroCare.
Paper Installation
- Push the printer cover release switch to the right to open the printer door of the
SpiroCare. Install recording paper with the side to be recorded appear on top.
Close the cover to finish paper installation process
.
Caution
Do not use SpiroCare in combination with any Electro-surgical equipment.
Caution
Users must use either the same type of mouthpiece or any other mouthpieces
with biocompatibility certification proven by international standards.
Содержание SpiroCare
Страница 1: ...SpiroCare Operation Manual Ver 1 1 2017 04 14 www ebionet com...
Страница 54: ...54 135 SCOMBA 1 1 SpiroCare Operation Manual FVC Test Pint Form...
Страница 57: ...57 135 SCOMBA 1 1 SpiroCare Operation Manual FVC Test Print Form...
Страница 62: ...62 135 SCOMBA 1 1 SpiroCare Operation Manual SVC Test Print Form...
Страница 67: ...67 135 SCOMBA 1 1 SpiroCare Operation Manual MVV Test Print Form...
Страница 72: ...72 135 SCOMBA 1 1 SpiroCare Operation Manual Calibration Print Form...
Страница 100: ...100 135 SCOMBA 1 1 SpiroCare Operation Manual Data Erase All data in Files and Worklist will be deleted...