FC1400 User Manual
FC1400OM-2.04 91 / 95
FC1400 Operation Manual
Acoustic Output Tables
MC65R1S
– Pulsed Doppler Mode
Index
MI
TIS
TIB
TIC
scan
Non-scan
Non-scan
Aaprt<= 1
Aaprt
> 1
Global Maximum : Index Value
0.0164842
-
0.00168143
-
0.0130577
0.00869565
IEC
FDA
Unit
Associated
Acoustic
Parameter
pra
pr.3
(MPa)
0.0164807
P
Wo
(mW)
-
0.4
0.4
0.4
min of
[P
α(zs),
I
ta,α(zs)]
min of
[(W.3(Z1),
ITA.3(z1)]
(mW)
-
zs
z1
(cm)
-
zbp
zbp
(cm)
-
zb
zsp
(cm)
1.8
1.8
z at max.
Ipi,
α
zsp
(cm)
deq(zb)
deq(zsp)
(cm)
1.0865
ƒawf
fc
(MHz)
0.999572
-
0.999572
-
0.999572
0.999572
Dim of
Aaprt
X
(cm)
-
0.4
-
0.4
0.4
Y
(cm)
-
0.4
-
0.4
0.4
Other
Information
td
PD
(μsec)
59.9647
prr
PRF
(Hz)
3906
pr at max.
Ipi
pr@PIImax
(MPa)
0.0175374
deq at
max. Ipi
deq@PIImax
(cm)
1.0865
Focal
Length
FLX
(cm)
-
2
-
2
FLY
(cm)
-
2
-
2
Ipa,
α at
max. MI
IPA.3@MImax
(W/cm^2)
0.00154839
Operating
Control
Conditions
Frequency
(MHz)
1.0
-
1.0
-
1.0
1.0