FC1400 User Manual
FC1400OM-2.04 45 / 95
FC1400 Operation Manual
2) TOCO screen
If the TOCO area on the screen is selected and touched, the following selection window pops up on
the screen:
①
Title
TOCO: Displays only uterine contraction but not fetal movements
TOCO+ FM: Displays both fetal movements and uterine contraction simultaneously
②
TOCO Zero value: Displays the reference value that appears when Zeroing has been done
③
TOCO: Displays the degrees of uterine contraction
3) Setup
If the TOCO area of the screen is touched, the following selection window pops up on the screen:
▣
Zeroing
This is used to adjust the zero point of TOCO. If this is touched, the TOCO value at the moment will
be set up as Offset value.
▣
Offset
This is used to set up a reference value to be used when adjusting the zero point of TOCO. One of the
values 0, 10, and 20 can be selected.
①
②
③