FC1400 User Manual
FC1400OM-2.04 13 / 95
FC1400 Operation Manual
General Precaution on Electric Safety
Check the items listed below before operating the equipment.
- Be sure that power supply line is appropriate to use. (100~240V AC)
- Be sure that the entire connection cable of the system is properly and firmly fixed.
- Be sure that the equipment is completely grounded. (If not, this might cause problems to occur in the
product.)
- Is there any damage -
which may affect the monitoring or safety of the patient ― to any of the
equipment or accessories?
NOTE
1.
The equipment should not be placed in the vicinity of electric generator, X-ray, broadcasting
apparatus to eliminate electric noise during operation. Otherwise, it may cause incorrect results.
Isolated power line is important for FC1400 To use same power source with other electric
instruments may cause incorrect results.
NOTE
FC1400 is classified as listed below ;
- Do not use the equipment in the vicinity of flammable anesthetics and solvents.
- The monitor is designed to fulfill safety requirements according to IEC 601-1/EN 60601-1
(Class I), Degree of protection against electrical shock type-BF.
NOTE
In hospitals, doctors and patients are exposed to the danger of uncontrollable electric currents.
These electric currents are generated by potential differences between the equipment and
conductive objects that may come into contact with the equipment. Please make sure that the
auxiliary equipment connected to the equipment to solve this problem satisfies EN60601-1-
1:1996.
Warning
Avoid contact with the patient during any defibrillation (may result in serious injury or
death). To avoid the danger of serious electric burning, shock, or injury, everybody
should stay away from the bed and should refrain from touching any equipment
connected to the patient.