FC1400 User Manual
FC1400OM-2.04 15 / 95
FC1400 Operation Manual
Maintenance and Washing Equipment
Connection
Although FC1400 and its accessories can be cleaned in many ways, please use the methods
recommended below to avoid unnecessary damage to or contamination of the equipment.
If any dangerous material other than those designated for cleaning has been used, the resulting
contaminated or damaged equipment will not be repaired free of charge even during the warranty
period.
Make sure that the monitor, probe, cable, and accessories are free of sand or dust. Carefully check
the equipment after each cleaning or disinfection. If degeneration or damage has been found, do not
use the equipment.
Please take note of the following:
- Be sure not to leave any cleaning/disinfecting chemical residue on the surface of the equipment.
After allowing sufficient time for the chemical to work, wipe off all residues with a cloth damp with
water.
- Prevent any fluid from seeping into the monitor, module, or accessories.
- The monitor, module, or accessories should not be soaked in any fluid; protect them from water
drops or prevent water from being splashed on them.
- Never use any abrasive material (steel wool or silver polish).
- Never use any bleaching agent.
- Do not use any kind of drying equipment such as heater, oven (including microwave oven), hair
dryer, or heating lamp.
NOTE
After cleaning the equipment, carefully check the main body and the probe.
Component cleaning Electric cable and lead wire
NOTE
- Do not use acetone or ketone solvents for cleaning
- Do not use an autoclave or steam cleaner.
- Do not mix the cleaner with any disinfecting solution (toxic gases may be produced).
- Please turn off the power before cleaning; do not pour water into the item being cleaned