FC1400 User Manual
FC1400OM-2.04 33 / 95
FC1400 Operation Manual
▣
Changes in the screen in relation to the state of battery connection
The battery condition should be displayed in relation to connection of Battery and AC power and
Battery condition as follow
* No Battery
: connect power to the equipment to use it
* Battery with little or no power
: 2 images are toggled
* Battery only
( Low),
( Medium),
( Full )
* AC and Battery
( Low),
( Medium )
( Full)
▣
Battery replacement
When the battery of this equipment is to be replaced, the same type of battery should be used.
-
Type : SDI1865L240 3S2PMXZ,Lithium Ion Rechargeable Battery Pack
(11.1V/4800mV)
-
Time: The battery will be automatically charged when the equipment is connected to AC power,
and the battery cannot be charged when it is separated from the equipment. The battery can be
charged at least 500 times; if using the battery for 20 minutes or longer after it is completely
charged is difficult, please replace the battery with a new one. If the battery has been damaged,
or battery fluid has leaked out, please replace the battery immediately. Do not use a damaged
battery in the equipment.
-
Method: Connect the connector as shown in the following image. (Connector cannot be plugged
inversely)
< Connect the Battery>