FC1400 User Manual
FC1400OM-2.04 83 / 95
FC1400 Operation Manual
The
FC1400
system is intended for use in the electromagnetic environment specified below.
The customer or the user of the
FC1400
system should assure that it is used in such an environment
Immunity test
IEC 60601
Test level
Compliance level E l e c t r o m a g n e t i c environment -
guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80
MHz
3 Vrms
150 kHz to 80 MH
z
Portable and mobile RF communications
equipment should be used no closer to an
y part of the system, including cables, tha
n the recommended separation distance c
alculated from the equation applicable to t
he frequency of the transmitter.
Recommended separation distance
Radiated RF
IEC 61000-4-3
3 V/m
80.0 MHz to 2.
5 GHz
3 V/m
80.0 MHz to 2.5 G
Hz
Recommended separation distance
Where
P
is the maximum output power rat
ing of the transmitter in watts (W) accordi
ng to the transmitter manufacturer and
d
i
s the recommended separation distance i
n meters (m).
Field strengths from fixed RF transmitters,
as deter-mined by an electromagnetic site
survey,
(a) Should be less than the compliance le
vel in each frequency range (b).
Interference may occur in the vicinity of
equipment marked with the following sym
bol:
Note 1)
U
т is the A.C. mains voltage prior to application of the test level.
Note 2)
At 80 MHz and 800 MHz, the higher frequency range applies.
Note 3)
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.