BM1VOM-1.24
8
.EtCO2 107
BM1VET Operation Manual
APNEA ALARM:
This performs a function to set the display of apnea message alarm.
This displays a “apnea” message at the center of parameter window as shown in the figure below
with apnea alarm on in case of apnea until the set apnea period is passed through.
RPM
A PENA
With apnea alarm off, measured values are displayed instead of message.
0
RPM
Et CO2
30
10
0
Fi CO2
0
Warning
If defibrillation is performed while doing CO2 monitoring, remove the CO2 FilterLine from animal
Getting in touch with sensor cable without removing the FilterLine can result in serious electrical
burn, shock, or injury due to electric discharge energy.
Note
In the following monitoring conditions, the measured values may be inaccurate. Read the
measured values carefully.
1. When using this in an environment using nitrous oxide gas of high concentration
2. When using this in an environment where abrupt temperature change takes place
3. When using this in an environment with severely high humidity.
Caution
⚫
The measured values may be inaccurate when using this equipment for animals who have
very fast or irregular respiration.
⚫
When measuring CO2 from the animal under the anesthesia, check it when gas mixture
comes in. Otherwise, the measured result values may be inaccurate.
⚫
When using a anesthesia machine that uses a volatile anesthetic, CO2 values may be
inaccurate.