BM1VOM-1.24
6.NIBP 156
BM1VET Operation Manual
Note
As the value of NIBP can vary according to the age and sex of an animal, the user needs to set
up right data in Parameter Menu before measurement.
The NIBP cable connector is insulated and it is defibrillator-proof(
). Use only the NIBP
cuffs listed in the enclosed publication.
WARNING
Noninvasive blood pressure monitoring is not recommended for animals with extremely high or
low heart rate or hypotension, hypertension, arrhythmias. The software algorithm cannot
accurately compute NIBP on animals with these conditions.
Note
As the value of NIBP can vary according to the age and sex of an animal, the user
needs to set up right data in parameter Menu before measurement. Tubes between the
cuff and the monitor must not be kinked or blocked.
The air pad should be exactly over the brachial artery. Tubing is immediately to the
right or left of the brachial artery to prevent kinking when elbow is bent.
The maintenance is performed every 2 years.
Check the following list to ensure device operates properly and safely at all times.
1. Check for proper cuff size.
2. Check for residual air left in the cuff from a previous measurement.
3. Make sure cuff is not too tight or too loose.
4. Make sure cuff and heart are at same level, otherwise hydrostatic pressure will offset the NIBP
value.
5. Minimize animal movement during measurement.
6. Watch for pulses paradox us.
7. Check for leak in cuff or tubing.
8. Animal may have a weak pulse.