CAPNOSTAT® 5 CO2 Sensor Adapter Zero
Adapter Zero
1. Set the BM1Vet to the zeroing function.
2. Connect the CAPNOSTAT 5 CO2 Sensor
3. Place the CAPNOSTAT 5 CO2 Sensor onto a clean and dry CO2 adapter that is exposed to
room air and away from all sources of CO2, including the ventilator, the animal
’s breath and
your own.
4. Start the adapter zero. The maximum time for a CAPNOSTAT zero is 40 seconds. The
typical time for a zero is 15~20 seconds.
Note
For best result, connect the CAPNOSTAT 5 CO2 Sensor to an adapter and wait 2 minutes
before performing the Adapter Zero procedure.
Zeroing the LoFlo Module
Note
A Sample Cell Zero is not required when switching from one sampling accessory
to another.
Sample Cell Zero
To perform a Sample Cell Zero:
1. Set the BM1Vet to the zeroing function.
2. Connect the LoFlo C5 CO2 Module and, if necessary, wait for the sensor warm-up message to
clear.
3. Connect a LoFlo Sampling accessory to the LoFlo C5 CO2 Module, and make certain that the
accessory is exposed to room air and away from all sources of CO2, including the ventilator, the
animal
’s breath and your own.
4. Start the Sample Cell Zero. The maximum time for a LoFlo zero is 40 seconds. The
typical time for a zero is 15-20 seconds.
Note
For best results, wait 5 minutes to allow the LoFlo C5 CO2 Module to warm up before
performing the Sample Cell Zero procedure.