BM1VOM-1.24
10. SPECIFICATION 181
BM1VET Operation Manual
SpO
2
capacity
· Saturation Range
:
0% to 100% oxygen proportion
· Pulse Rate Range
:
30 to 254 bpm
· SpO
2
accuracy
:
70% to 100% ±2 digits, 0% to 69% unspecified
· pulse accuracy
:
±2 bpm
· Sensor
Red 660nm, 2mW (typical)
Infrared 905nm, 2-2.4mW (typical)
· Minimum Signal:
0.05% modulation (Low
perfusion level performance and
Amplitude
limitation validation using
FLUKE Index 2 Oximetry Simulator)
NIBP capacity
· Technique
:
Oscillometric
· Measurement mode:
- Manual
:
Single Measurement
- Auto
:
automatic Intervals of 1MIN, 2, 3, 4, 5, 10, 15, 20, 30, 1HR, 2, 4, 8
· Pressure Display
:
0 to 300 mmHg
· Blood Pressure Measurement Range: ( ADULT / PEDIATRIC / NEONATE )
- systolic
:
40 to 250 mmHg / 40 to 230 / 40 to 120
- Mean Arterial Pressure
:
30 to 220 mmHg / 30 to 190 / 30 to 100
- Diastolic
:
20 to 200 mmHg / 20 to 170 / 20 to 90
Temperature Unit Performance Specifications
· Range
:
10°C to 40°C (50°F to 104°F )
· Accuracy
:
25°C to 45°C ± 0.1°C, 15°C to 24°C±0.2°C
· Sensor :
YSI 400 Series compatibility
EtCO2 Module Performance Specifications
· EtCO2 Range
:
0 to 150 mmHg , 0 to 19.7% , 0 to 20 kPa
· EtCO2 Accuracy
:
0 to 40 mmhg : ± 2mmHg
41 to 70 mmhg : ± 5%
71 to 100 mmhg : ± 8%
101 to 150 mmhg : ± 10%
· Respiration Range
:
0 to 150 bpm ( breath per minute )
· Respiration Accuracy
:
± 1bpm
Accessories Included:
· NIBP tubing, 3m long
1 EA
· Infant cuff, 25-35 Cm
1 EA
· SpO
2
extension cable 2m
1 EA
· SpO
2
sensor
1 EA
· DC adapter, 15VDC, 2.0A (JMW128 Made in AULT Co., Ltd.)
1 EA
Option
· Temperature sensor (skin)
· EtCO2 Module and accessories ( Refer to the EtCO2 chapter of this manual for details)