BM1VOM-1.24
6. NIBP 80
BM1VET Operation Manual
ALARM LIMIT
Alarm setting Numeric Value of Systolic, Diastolic, and mean pressure is 10 ~ 300mmHg.
1. Move the
▶
mark to select one from RETURN, NIBP-S, NIBP-M, or NIBP-D, and press.
2. Press the key at NIBP-S, and move to LOW, and press again.(The user gets the same result
regardless of the LOW-HIGH, or HIGH-LOW order.)
3. When the color has changed, move it again to select a target value, and press.
4. Press the key at HIGH. When the color has changed, move to the right to select a target value,
and press.
5. Set up or revise the values of NIBP-M and NIBP in the same way as above.
6. With the selection of RETURN, the user can get out of the window.
ALARM
LIMIT
ALARM
SOUND:
ON
PREV
MENU
MAIN
MENU
NIBP ALARM LIMIT
RETURN
UNITS
LOW
HIGH
NIBP-S
mmHg
80
200
NIBP-M
mmHg
40
140
NIBP-D
mmHg
20
120
Содержание BM1VET
Страница 53: ...BM1VOM 1 24 52 BM1VET Operation Manual 3 SETUP 3 1 SETUP DISPLAY DEMO USER SERVICE MAKER SERVICE ...
Страница 61: ...BM1VOM 1 24 60 BM1VET Operation Manual 4 TREND 4 1 TREND GRAPHIC TREND TABULAR TREND ...
Страница 77: ...BM1VOM 1 24 6 NIBP 76 BM1VET Operation Manual CAT CUFF Placement DOG CUFF Placement ...