BM1VOM-1.24
8
.EtCO2 103
BM1VET Operation Manual
- BAROMETRIC PRESSURE:
This setting is used to set current Barometric Pressure.
- GAS TEMPERATURE:
This setting is used to set temperature of the gas mixture. This
setting is useful when bench testing using static gasses where
the temperature is often room temperature or below.
-NO BREATH DETECT TIMEOUT:
This setting is used to set the no breaths detected time-out. This
time-out is the time period in seconds following the last detected
breath at which the Capnostat will signal no breaths detected.
-O2 COMPENSATION: O2 Compensation
-ANESTHETIC AGENT: Anesthetic agent
-BALANCE GAS:
Use this setting to correct for the compensation of the gas
mixture administered to the animal. Anesthetic agent is ignored
when the balance gas is set to helium.
-CURRENT ETCO2 TIME PERIOD: This setting is used to set the calculation period of the ETCO
2
value. The end-tidal CO
2
value is the highest peak CO
2
value of
all end of expirations (end of breaths) over the selected time
period. If less than two breaths exist in the selected time period,
the value will be the maximum ETCO
2
value for the last two
breaths.
-CURRENT CO2 UNIT:
Continuous waveform mode commands (the CO
2
Waveform
Mode command [command 80h] and the CO
2
/O
2
Waveform
Mode command [command 90h] ) MUST NOT be active when
this command is used otherwise this command will be ignored
and the setting will remain unchanged.
-SLEEP MODE:
Sleep mode is used to save power when the host monitor is in
standby mode. There are two sleep modes available for the
Capnostat. Using Sleep Mode 1 maintains the heaters so the
Capnostat is able to run immediately after exiting the sleep
mode. Mode 2 will require the Capnostat to go through its warm
up sequence when exiting this mode and a delay will be
introduced until the system has stabilized.
-ZERO GAS TYPE:
When performing a zero on room air, this setting should be set
to room air (the default). Only change to nitrogen (N
2
) when
performing a zero on 100% N
2
gas; this is provided for use in a
laboratory environment.
-DISABLE SAMPLING PUMP:
This setting allows the pump to be forced off. In Normal
Operating Mode, the pump will be turned on when the sampling
cell is connected and no pneumatic system errors are detected.
In Pump Disabled Mode, the pump will remain off in all
circumstances.