5.4 TROUBLE SHOOTING
PROBE OFF Condition
When using a reusable finger probe, there is a system alarm to alert you when the probe is off the
Monitor
. The monitor defaults this “ PROBE OFF” condition as a System Warning alarm. You can,
however, set it as a System ALARM LEVEL in Monitor Defaults.
SPO2 Messages
Below is a list of system status alarm messages which may be displayed in the SPO2 parameter
window during monitoring.
CHECK PROBE
Reusable finger probe is off the animal. Check the probe.
The factory default for this alarm is
MESSAGE ALARM.
PULSE SEARCH
Detection by the monitor of a repeatable pulse has ceased. Check the animal and the probe site.
POOR SIGNAL
The SPO2 signal is too low. No SPO2 data is displayed. This can be due to a low animal pulse,
animal motion, or some other interference. Check the animal and the probe.
LOST SIGNAL
SPO2 data continues to be displayed, but the quality of the signal is questionable. Check the animal
and the probe.
ARTIFACT
The SPO2 signal is
animal's motion artifact and noise
No SpO2 data is displayed. One of the following conditions is indicated:
⚫
defective or damaged probe,
⚫
defective or damaged cable
⚫
probe is off the animal, or
⚫
Detection of a repeatable pulse has ceased.
⚫
Check the probe and cable: reposition or replace as needed.