BM1VOM-1.24
3. SAVE RECORD 134
BM1VET Operation Manual
3.4 Measure with Spot Mode
Measure after setup a mode to MANUAL.
MAKER
SERVICE
SYSTEM
KEY
SOUND:
ON
DISPLAY
SAVE
MODE:
MANUAL
USER
SERVICE
MAIN
MENU
It saves as press the button after measurement.
NIBP ending spot numeric value stores when NIBP is INTERVAL mode.
NIBP ending spot numeric value stores when NIBP is STAT mode
When NIBP is MANUAL mode, it saves measured numeric value after 60seconds of event below.
Event:
Input animal information
Measure NIBP
Measure SpO2
When new events occur in 60 seconds after previous event then it saves after 60 seconds of new
events occur.
All measured parameters are removed from the screen after finished with saving.
Search from record list to confirm measured result.
After saving, the animal ID initializes as UNKNOWN.
After saving, adjusted alarm limit numeric value becomes Default numeric value.