BM1VOM-1.24
8.TEMPERATURE 175
BM1VET Operation Manual
8.4 TROUBLESHOOTING
Check list
4. The temperature probe(YSI 400 series) is correctly positioned on the animal.
5. Temperature cable is attached to the monitor.
6. Temperature setup is adjusted, if necessary. Follow detailed procedures within this chapter.
TEMP Message
If you experience some problems with temperature monitoring, one of the following messages may
be displayed in the TEMP parameter window.
⚫
PROBE OFF : Probe is not properly connected. Check the probe.
- No temperature value will be displayed . Service on the monitor is required.
Warning
To measure the peripheral temperature, attach the probe to the ankle or palm.
If the patient sweats heavily or moves violently, fasten the pad with surgical tape.
NOTE
When the measuring site is exposed directly to air, the temperature may be lower than normal.
It take about 20 to 30 minutes to reach the equilibrium temperature after attaching the sensor.
Содержание BM1VET
Страница 53: ...BM1VOM 1 24 52 BM1VET Operation Manual 3 SETUP 3 1 SETUP DISPLAY DEMO USER SERVICE MAKER SERVICE ...
Страница 61: ...BM1VOM 1 24 60 BM1VET Operation Manual 4 TREND 4 1 TREND GRAPHIC TREND TABULAR TREND ...
Страница 77: ...BM1VOM 1 24 6 NIBP 76 BM1VET Operation Manual CAT CUFF Placement DOG CUFF Placement ...