CL-10 Plus – User Manual
9 / 75
ECPUS vers 0.2 eng
2.3.5 Hydraulic
connections.
1. Hook the peristaltic tubes with lower diameter in position 6 (and 5, if required by the
method), without twisting them.
Warning: if tubing’s are twisted, analytical reliability is compromised. To be able to verify the correct tubing’s
assessment, mark the thick ends of each tube before hooking it, on the same side along the fusion line and
check that, once hooked, it remains marked on the same side.
2. Hook the 4 peristaltic tubes, with higher diameter, in position 1,2,3 and 4 without
twisting them. Careful that the ’L’ connections are not pushed to the far end of the
silicon but only halfway.
3. Place the bottle in the waste position.
4
5
6
1
2
3
TAMPONE/BUFFER
SCARICO/WASTE
ENZIMI/STARTER
7
8
9
Fig. 2