Sample
PHASE 2
: both electrode pumps n° 1 and n° 2 rotate, taking approximately half of the total
volume of the mixing chamber, filling both the electrodes with milk.
E -E
2
1
D1
1
2
At this stage the instrument reads, for 5 seconds, the pH difference between EL 2 - EL 1 =
∆
mpH; called D1. The electrodes contain exactly the same solution, the pH difference should
be zero (in fact between
±
150 mpH). The signal measured will be subtracted during the
following measure.
Example
: during the measurement the obtained
∆
mpH is 20, the following graph
∆
mpH/time
will be displayed
Pump
1
Pump 2
∆
mpH
t (s)
20
5
D1