CL-10 Plus – User Manual
75 / 75
ECPUS vers 0.2 eng
1 - Specifications were obtained by a gravimetric method with the temperature stabilised
between 21 and 22 °C. The values given take into account all variables, error included, both
normal warning of the handle, and the changing of the capillaries and pistons.
2 - Repeatability defined by dispensing 30 measures with distilled water, with a single
instrument fitted with Gilson capillaries and pistons.
7.5
How to Inject The Sample.
- Check the volume set on pipette.
- Aspirate slowly the sample taking care not to have air bubbles into the capillary.
- Wipe with laboratory paper any fluid outside the capillary.
- Insert the capillary into the CL10 measuring chamber, until gently touching the bottom and
push with moderate speed the push button. Do not release the push button before having
completely lifted the pipette from the measuring chamber.
ATTENTION:
- AVOID INSERTING THE CAPILLARY FORCEFULLY INTO THE MEASURING CHAMBER.
IT COULD DAMAGE OR AFFECT THE PIPETTING PERFORMANCES.
- REPLACE THE CAPILLARY AND PISTON IF YOU NOTICE SOME LIQUID INTO THE
CAPILLARY, BEYOND THE PISTON TIP.
-
AFTER EVERY INJECTION WIPE THE CAPILLARY, OTHERWISE YOU MAY
CONTAMINATE THE NEXT SAMPLE.