2.5.3 Electrodes
regeneration.
NOTE: Regenerating procedure of measuring electrodes is used to reactivate the instrument after a period of
halt (installation or after a period of rest); electrodes, after this procedure require a “stabilization” time; it is
therefore recommended to operate the Rest Mode procedure immediately after, using wash solution as a buffer.
Electrodes regenerating procedure:
1.
Reconstitute the wash solution, according to the instructions given in the kit.
2.
Place the reconstituted solution in
buffer
position and an empty bottle in
waste
position
(Fig. 2, § 2.3.5).
3.
Run 3
Clean
cycles.
4.
Replace the bottle containing wash solution with a bottle of
Regenerating
Solution
in
buffer position.
5.
Run 3
Clean
cycles.
6.
Wait for 5 minutes.
7.
Replace the
Regenerating
Solution
with a bottle of distilled water and run 2
Clean
cycles.
8.
Replace the bottle of distilled water with reconstituted wash solution, again and run 3
Clean
cycles.
9.
Empty the waste bottle.
2.5.4
Standby mode (Rest)
For the CL-10 Plus instrument there are two available different
standby mode
(REST mode):
1
From the computer
,
REST
mode
program ( § 2.3.4).
From
Service
window enter
Enter Rest Mode
command
Switch the monitor off
Leave the PC on
Leave the CL-10 Plus instrument on
ATTENTION: Check if there is enough Wash Solution for all the standby period (the washing cycle automatically
runs every 120 minutes and consumes from 4 to 9 mL of wash solution each cycle, depending on the execution
method selected).
2 From the instrument
, REST mode: the instrument stays at 37°C and the time laps
between consecutive clean cycles is 90 minutes.
Switch the PC monitor off.
Switch the CL10 Plus off.
Allow 5 seconds, then switch the CL10 Plus on again.
The instrument runs 1 clean cycle every 90 minutes at 37°C.
ATTENTION: Check if there is enough Wash Solution for all the standby period (the washing cycle automatically
runs every 90 minutes and consumes from 4 to 9 mL of wash solution each cycle).
ALLOW STANDBY OVERNIGHT, at least; it is not required for 2-3 hours interruption of working session